Problem

In order for your 1PPS signal to be recognised by Paragon-X, you need to make sure that your input signals (both reference and measurement) are at the correct level.

...

Panel
bgColor#fff

Front panel 1PPS (aux) input voltage

The maximum 1PPS input voltage which can be applied to the Paragon-X upper aux measurement port depends on its duty cycle. The table below specifies the maximum upper voltage level assuming the V(lo)=0:

Duty cyclemax V(Hi)
0.16.3
0.24.5
0.33.65
0.43.2
0.52.8
0.62.6
0.72.4
0.82.24
0.92.36

So, for example, a 1PPS signal with a 50:50 duty cycle should have a maximum V(Hi) of 2.8V. Application of a signal voltage higher than those in the table may cause damage to your Paragon-X hardware and so must be avoided.

For optimal detection of your 1PPS pulse, the threshold should be set in the Measurement Ports tab of the Setup Interface dialogue to a value approximately half of V(Hi).

1PPS signal timing

The measurement 1PPS signal must be received within 500ms of the reference 1PPS input. Signals outside this window will be indicated as a missing pulse in the Paragon-X GUI and CAT by a vertical red line:

1PPS reference input

The threshold value for your 1PPS reference input signal (connected to the rear of Paragon-X) may also be adjusted, via the References tab of the Setup Interface dialogue:


Related articles

Filter by label (Content by label)
showLabelsfalse
max5
spacescom.atlassian.confluence.content.render.xhtml.model.resource.identifiers.SpaceResourceIdentifier@957
showSpacefalse
sorttitle
typepage
cqllabel in ( "paragon-x" , "1pps" ) and type = "page" and space = "KB"
labelsparagon-100g paragon-neo paragon-x

...