A PRTC test with Paragon-neo is typically connected like this:
...
Info |
---|
The Opt 133 converter can be removed from the 1PPS reference path if Paragon-neo is running R11 and later software. |
...
Compensation Settings
Follow steps 1-5 to compensate the PRTC and Paragon-neo:
The PRTC must be compensated for the RF cable delay from the GNSS simulator - how this is done is device specific - it will typically be entered in nanoseconds of delay, or meters of cable.
Some PRTC’s may allow for output signal compensation, it is recommended that initially the PRTC is tested with zero output signal compensation, and once this has been validated then the operation of any output compensation can be tested.
Paragon-neo
...
Paragon-neo requires the following compensation set when testing a PRTC:
1PPS reference cable 1PPS Reference Cable compensation accounts for the difference between propagation delays of the simulator 1PPS output to the Paragon-neo reference input. In the example above this is 10+18+10 = 38ns.
Because the PRTC RF input has been compensated (step 1), the Paragon-neo 1PPS Reference Cable compensation number will be a positive number.If the PRTC has a PTP output, the Paragon-neo DUT Ethernet cable compensation accounts for the propagation delay of the cable between the PRTC PTP output and Paragon-neo measurement port. In the example above this is 20ns.
Paragon-neo 1PPS measurement cable compensation accounts for the propagation delay of the cable between the PRTC 1PPS output and Paragon-neo measurement port. In the example above this is 10+18+10 = 15ns.
The Using the example propagation delays from the pictures above, compensation values are entered into the Paragon-neo GUI through two GUI application panels.
The 1PPS reference compensation is compensated in the PTP Emulation > Test Configuration panel.
The Ethernet cable between PRTC and Paragon-neo is also compensated in the Test Configuration panel
...
.
...
The 1PPS measurement cable
...
is compensated in the 1PPS Time Error Measurement + Time of Day panel.
The 1PPS reference cable compensation can also be entered in this panel, or it will be copied from the PTP Emulation Test Configuration panel.
...
The PRTC and Paragon-neo is are now compensated and ready to run Time Error tests on the PRTC.
Info |
---|
Please also refer to the following related FAQ articles for further information: Paragon-neo Option 133: how to measure 1PPS propagation delays |
...