Panel |
---|
Timebase Reference statusInternal Rubidium is used as Timebase Reference External signal is used as Timebase Reference Rubidium statusRubidium is ready Rubidium is warming up Rubidium is not ready or Rubidium malfunction detected Rubidium disciplining statusRubidium is disciplining and locked with a phase offset less than +/-100ns Rubidium is disciplining and locked with a phase offset greater than +/-100ns Disciplining is off. Either disciplining mode is set to “Never” or disciplining mode is set to “Not during measurement” and measurement is currently in progress or GPS is the disciplining source and GNSS is not locked to at least three satellites. Rubidium is disciplining and acquiring phase lock Disciplining is turned on, but no valid disciplining signal is present Ext 1PPS statusExternal 1PPS is present and is generating a valid 1PPS signal. Icon turns clear when Rb is in manual hold-over mode. External 1PPS is either not present or is not generating a valid 1PPS signal. Icon turns clear when Rb is in manual hold-over mode. GNSS statusGNSS malfunction GNSS is locked to at least 3 satellites and site survey is complete. GNSS is not locked yet, but some satellites are visible. No antenna or no visible satellites. Battery statusBattery is not installed Battery is installed and charging (charge percentage is indicated next to battery icon) Battery is installed and not charging. Battery will not charge when 100% charged, or while Rubidium is warming up or if the battery temperature is less than 0oC or greater than 45oC Battery is installed, error has occurred reading battery controller or a battery fault has been detected |