Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 2 Next »

A PRTC test with Paragon-neo is typically connected like this:

The blue lines show the cables connecting the GNSS Simulator PRTC, Paragon-neo and Opt 133 converter. The red numbers show the delays associated with each cable segment and the Opt 133 converter. If the PRTC is perfectly calibrated, it will have zero delay (i.e. zero time error) between the incoming GNSS signal and its 1PPS and PTP outputs.

Note: the numbers shown in this diagram are chosen to simplify the example. Please use precise cable and Opt 133 delays when calculating compensation numbers for a real test bench.

The Opt 133 converter can be removed from the 1PPS reference path if Paragon-neo is running R11 and later software.

Paragon-neo requires three compensation figures when testing a PRTC:

  1. 1PPS reference compensation accounts for the difference between propagation delays of the simulator RF output to PRTC and simulator 1PPS output to Paragon-neo reference input.

  2. 1PPS measurement cable compensation accounts for the propagation delay of the cable between the PRTC 1PPS output and Paragon-neo measurement port.

  3. Ethernet cable compensation accounts for the propagation delay of the cable between the PRTC PTP output and Paragon-neo measurement port.

The next diagram shows a timeline for example test network above. The simulators RF output and 1PPS output are precisely aligned. The upper timeline shows the RF signal propagation from the simulator to the PRTC and the PRTC 1PPS output propagation to Paragon-neo measurement input. The lower timeline shows the 1PPS signal propagation from the simulator to Paragon-neo reference input through cables and the Opt 133 converter.

The difference in the two signal paths must be compensated so that Paragon-neo can accurately measure performance of the PRTC.

In this example the 1PPS reference signal arrives at Paragon-neo before the RF signal arrives at the PRTC. Therefore we need a negative compensation value of 38 - 80 = -42ns to compensate the 1PPS reference signal. Effectively we are delaying the reference 1PPS.

If the propagation delay from the simulator to the PRTC is less that then 1PPS propagation delay from the simulator to Paragon-neo, the compensation would need to be a positive value. Effectively we are advancing the reference 1PPS.

The compensation values are entered into the Paragon-neo through two GUI application panels.

  1. The 1PPS reference compensation (-42ns in this example) is compensated in the PTP Emulation Test Configuration panel.
    The Ethernet cable between PRTC and Paragon-neo is also compensated in the Test Configuration panel (20ns in this example).

  1. The 1PPS measurement cable (15ns in this example) is compensated in the 1PPS Time Error Measurement panel.
    The 1PPS reference cable compensation can also be entered in this panel, or it will be copied from the PTP Emulation Test Configuration panel.

Paragon-neo is now compensated and ready to run Time Error tests on the PRTC.

Please also refer to the following related FAQ articles for further information:

Paragon-neo Option 133: how to measure 1PPS propagation delays

Paragon-neo: 1PPS Input Accuracy

  • No labels