I am trying to connect to the Paragon-X from my script. I get an error message that says that the connection has failed because the port is already in use.
First of all check that there are no Calnex programs running - Paragon-X GUI, CAT or PFV. Next check Task Manager for any Calnex processes and, if there are any, kill them. If the port is still locked, it is possible that it is held by an orphaned process. Try the following:
netstat -ano | findstr :9000
wmic process where (ParentProcessId=<PID>) get Caption,ProcessId where <PID> is the PID from above (in this case 31076)
Note: The above is a real example. It looks as though Wireshark also uses port 9000. Closing Wireshark allowed Paragon-X to be used remotely once more… |
Related articles appear here based on the labels you select. Click to edit the macro and add or change labels.
|